Product Features | ||
Medical & I.T.E. safety approvals | ||
2MOPP input to output isolation | ||
Leakage current≤ 100μA | ||
DOE efficiency level VI | ||
Standby power ≤0.3W | ||
12V to 24V outputs, up to 36W | ||
Interchangeable AC plugs | ||
Technical Data | ||
Input Voltage Range:90~264VAC | ||
Input Current:0.9A at 90VAC | ||
Output voltage Tolerance:±5% | ||
Environmental Requirements | ||
Operating Temperature: 0~45℃ | ||
Storage Temperature: -20~60℃ | ||
Operating Humidity:10~90%RH, non-condensing | ||
Storage Humidity::5~90%RH | ||
Operating Altitude: 5,000 M | ||
Safety Approvals | ||
Safety aAgency/Mark | Medical | I.T.E |
CB | IEC60601-1 | IEC60950-1 |
UL | ANSI/AAMI ES60601-1 | - |
CAN/CSA C22.2 NO.60601-1 | - | |
TUV | EN60601-1 | - |
RCM | IEC60601-1,Cert:SAA150158 | - |
CCC | - | GB4943.1 |
Mechanical Details | ||
102(L)*53(W)*36(H)mm | ||
Applications: | ||
Medical Device and ITE Equipment |
Models & Ratings | |||||||
Model No. | Voltage (V) | Current (A) | Ripple & Noise (Max: mV) PK-PK | Voltage tolerance | Line & Load Regulation | Efficiency (Average) Min | Start Up delay |
UE36LCP1-XXXYYYSPA | 12.0 | 0.01-3.00 | 200 | ±5% | 84.78% | ≤3S | |
12.1-13.0 | 0.01-2.67 | 200 | ±5% | 84.55% | ≤3S | ||
13.1-14.0 | 0.01-2.57 | 200 | ±5% | 84.78% | ≤3S | ||
14.1-15.0 | 0.01-2.40 | 200 | ±5% | 84.78% | ≤3S | ||
18.0 | 0.01-2.00 | 200 | ±5% | Line:±1% | 84.78% | ≤3S | |
18.1-19.0 | 0.01-1.89 | 200 | ±5% | Load: ±5% | 84.77% | ≤3S | |
19.1-20.0 | 0.01-1.80 | 200 | ±5% | 84.78% | ≤3S | ||
20.1-21.0 | 0.01-1.71 | 200 | ±5% | 84.77% | ≤3S | ||
21.1-22.0 | 0.01-1.63 | 200 | ±5% | 84.76% | ≤3S | ||
22.1-23.0 | 0.01-1.56 | 200 | ±5% | 84.76% | ≤3S | ||
23.1-24.0 | 0.01-1.50 | 200 | ±5% | 84.78% | ≤3S |